27
Display shows…
Error 1
Error 2
Error 3
Troubleshooting
Troubleshooting
In response to various environmental conditions, the generator will display
an error message. Understanding these messages will help you quickly
troubleshoot the problem. If the suggested actions below do not resolve the
issue, contact Electromed or your service provider.
Problem is…
The motor temperature
is above the maximum
acceptable limit. The
unit will NOT start in
this state, as damage
may occur.
The motor temperature
is below 50 degrees
Fahrenheit (10 degrees
Celsius) and needs to
warm up.
Motor rotation has not
been detected.
Take this action…
If the system is sitting in a
warm location, move it to a
cooler location and try again
later. If this does not resolve
the issue, contact Electromed
or your service provider.
Leave the unit plugged in
and turned on so the fan
can assist in warming up the
unit. With the unit power on,
the device will periodically
check the motor temperature
and after warming up, it will
automatically proceed to the
Home screen.
If this does not resolve the
issue, contact Electromed or
your service provider.
Contact Electromed or your
service provider.