equipped with two independent adjustable alarms (micro switches) which can
be set for low and high flow alarm. The readout has a scale, which shows you
the ordered flow range with a multiplier as a standard. The standard scales
have the following designations; for S2/S02 the scale goes from the number 4
to 8 (1:2 turndown). This is the value that you have to multiply the small mul-
tiplier (sticker) figure at the bottom of the scale with, in order to read the ac-
tual flow in the values you have ordered i.e. liters/min, m3/hours, USGPM.
etc. (also shown at the bottom of the scale) For the S25/S05, the scale goes
from 1 to 5 and hence, this is the value you use together with the multiplier at
the bottom of the scale, to read the actual Flow through the Monitor.
The Control Unit can also be fitted with a special ordered scale (direct reading
scales) as an option, for example; l/min, m3/h, USG/min etc. for easy reading
and then you have no multiplier at the bottom of the scale.
Specifications
The only flow difference between the S2/S02 and S25/S05 is the turn down of
the flow range i.e. the S2/S02 has a 1:2 turn down (for example; 50 –100
l/min) and the S25/S05 has a turn down of 1:5 (for example; 20 – 100 l/min).
Accuracy: <+/–5% F.S (full scale) within 20-80 % of the chosen Flow range
<+/–10% F.S.(full scale) within 100% of the chosen Flow range
The accuracy stated is achievable if the installation instruction
is followed given in this manual. It is recommended that you
always choose the Flow Range of the Flow Monitor so that the
normal flow is in the middle of the Monitor Flow Range.
Make sure that the expected alarm set points are within the
chosen flow range. For example: If you have a flow of 110
l/min maximum and the normal Flow is at 90 l/min, chose the
Eletta Flow Monitor S2 with a Flow Range of 60–120 l/min.
This will give you the highest accuracy since your flow is in
the middle of the Monitor Flow Range and give you a lower
pressure loss.
Repeatability: < 2 %
Pressure: Max:16 bar, (232 PSI), higher test pressure as an option.
Min: A line pressure of apx.: 0,7 – 1,0 bar (10 – 14,5 PSI) is
required for proper operation.
Temperature: Control Unit
–20°C to 90°C (–4°F to 200°F), standard
1
1.2
1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eletta S2 and is the answer not in the manual?

Eletta S2 Specifications

General IconGeneral
BrandEletta
ModelS2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals