4
Blood Glucose Monitoring System
Limitations
Inaccurate results may occur in
• Severely hypotensive individuals
• Patients in shock
• In a hyperglycemic-hyperosmolar state with or without ketosis
Do not use on critically ill patients.
Do not use on neonates.
Do not use the system above 10,335 feet (3,150 meters) in altitude.
Do not use if hematocrit exceeds the acceptable range between 10% to 70% when testing
blood glucose.
Severe dehydration (excessive water loss) may cause inaccurate results.
For In Vitro Diagnostic only.
For Over the Counter use.
Perform an Alternative Site Testing (AST) only if your glucose level is not changing
rapidly.
For single-patient use only.
Do not use AST measurements to calibrate Continuous Glucose Monitors (CGMs).
Do not use AST measurements for insulin dosing calculations.