5
Blood Glucose Monitoring System
The following drugs do not generally interfere with the Embrace
®
WAVE+ Blood Glucose
Monitoring System at normal or therapeutic levels. However, higher concentrations (listed
below) were found to interfere with glucose measurements.
If you are taking acetaminophen or acetaminophen containing drugs (for example
Tylenol; at blood concentrations > 10 mg/dL) you may get inaccurate results with this
system. If you are unsure, then please contact your doctor.
If you are taking Paralidoxime Iodide (PAM) or Paralidoxime Iodide (PAM) containing
drugs at blood concentrations > 50 mg/dL, you may get inaccurate results with this
system. If you are unsure, then please contact your doctor.
If you have certain conditions that may cause your blood level of uric acid to rise (> 15
mg/dL in your blood), such as gout or kidney disease, then your blood glucose results
may be inaccurate with this meter. If you are unsure, then please contact your doctor.
Do not use during or soon after xylose absorption testing since xylose may cause
inaccurate glucose results. Ask your doctor how long to wait before performing a glucose
test.