Maintenance (continued)Table 2-2:
Daniel MeterLink utility Diagnostics Action(s)
Meter Monitor (Detailed) view Flow Profile Flow profile velocity for me-
ters with four Chordal paths
- Chords A and D shown on
the Bar graph should be of
equal lengths and at 5 ft/s
should be 0.89 and Chords
B and C should be 1.0 indi-
cating a good flow profile.
If the velocity range is
greater than a 10% differen-
tial between chords, a deg-
radation in the symmetry is
indicated by a change in
the flow velocity for the
chord and the measured
SOS. A chord that causes
the spread for the Speed of
Sound to vary more than
0.35% of the Average
Speed of Sound, the bar for
that chord turns yellow.
Clicking one of the arrows
at the top left of the graph
will change the chart to
SOS Differences from the
average meter speed of
sound. This provides a
quick indication on how
much spread in speed of
sound is between the
chords. Negative values are
shown in blue.
Compare Gains and Signal
to Noise (SNR) ratios deci-
bel values with the values in
the Maintenance log In-
spection report from the
meter flow calibration.
Check chord average signal
amplitudes with the meter
baseline values in the Main-
tenance log Inspection re-
port.
The meter may not be in
measurement mode or
there are too few operating
chords.
Troubleshooting
Maintenance and Troubleshooting manual 29