Micro Motion
®
Model D and DT Sensors Instruction Manual 31
Wiring continued
Model D600 wiring to Model 3700
Model D600 wiring to RFT9739 field-mount transmitter
Model D600 wiring to RFT9739 rack-mount transmitter
Model 3700
terminals
Red
Brown
Ye l l o w
Black (drains)
Violet
Orange
Green
White
Blue
Gray
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Yellow
Clip drain wire back
Flowmeter
cable
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Ye l l o w
Black
(Drains from all
wire sets)
Maximum cable length 1000 ft. (300 m)
Green
White
Brown
Violet
Yellow
Orange
Blue
Gray
Red
Model D600 sensor
junction box
terminals
Prepare cable in accordance with the
instructions that are shipped with the cable.
Do not allow shields to contact sensor
junction box.
Green
White
Brown
Violet
Ye l l o w
Orange
Blue
Gray
Red
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Ye l l o w
Clip drain wire back
Model D600 sensor
junction box
terminals
Flowmeter
cable
Field-mount
RFT9739 terminals
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Brown
Orange
Green
Violet
White
Gray
Blue
Yellow
Red
Black (Drains)
Maximum cable length 1000 ft. (300 m)
Prepare cable in accordance with the
instructions that are shipped with the cable.
Do not allow shields to contact sensor
junction box.
Rack-mount
RFT9739 terminals
Red
Ye l l o w
Orange
White
Gray
Brown
Black (Drains)
Violet
Green
Blue
B2
B4
B6
B8
B10
Z2
Z4
Z6
Z8
Z10
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Ye l l o w
Clip drain wire back
Flowmeter
cable
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Maximum cable length 1000 ft. (300 m)
Green
White
Brown
Violet
Ye l l o w
Orange
Blue
Gray
Red
Model D600 sensor
junction box
terminals
Prepare cable in accordance with the
instructions that are shipped with the cable.
Do not allow shields to contact sensor
junction box.

Other manuals for Emerson Micro Motion

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion and is the answer not in the manual?

Emerson Micro Motion Specifications

General IconGeneral
CategoryMeasuring Instruments
ManufacturerEmerson
BrandMicro Motion
TypeCoriolis Flow Meter
Output4-20 mA, HART, Foundation Fieldbus, Profibus, Modbus
Power Supply24 V DC
MaterialStainless Steel, Hastelloy
Process ConnectionFlanged, Threaded
CertificationsATEX, IECEx, CSA, SIL
Pressure RatingVaries by model; up to 10, 000 psi (690 bar)
Temperature RangeVaries by model; -400 to 750°F (-240 to 400°C)