8
Safety Manual
00809-0400-5100, Rev AB
Installation and Configuration
November 2019
Installation and Configuration
2.2 Installation in SIS applications
The Rosemount
™
5900 Radar Level Gauge and Rosemount 2410 Tank Hub should be
installed and configured as described in the reference manual. The materials must be
compatible with process conditions and process fluids. No special installation is required in
addition to the standard installation practices outlined in the reference manuals:
Rosemount 2410 Tank Hub Reference Manual
(Document No. 00809-0100-2410)
Rosemount 5900S Radar Level Gauge Reference Manual
(Document No. 00809-0100-5900)
Rosemount 5900C Radar Level Gauge Reference Manual
(Document No. 00809-0100-5901)
Rosemount Tank Gauging System Configuration Manual
(Document No. 00809-0300-5100)
Note
Installation drawings must be considered for installation of devices in a Rosemount Tank
Gauging Safety System.
Note
The Rosemount 5900 Radar Level Gauge and Rosemount 2410 Tank Hub are not
safety-rated during maintenance work, configuration changes, or other activity that affects
the Safety Function. Alternative means should be used to ensure process safety during such
activities.