Safety Manual
00809-0400-5100, Rev AB
Operation and Maintenance
November 2019
Operation and Maintenance
3.5.2 High level alarm test with simulated reflector
Prior to running a proof test you will have to ensure that the Simulated Reference
Reflector is properly calibrated and configured
(1)
. Ensure that High Alarm is set to an
appropriate level below the Proof Test Reference Reflector.
The High Alarm Test with simulated reference reflector should be combined with:
One-point level proof test
Output verification test such as Analog Output or Relays K1/K2.
To run a proof test for a Rosemount 5900 with simulated reference reflector do the
following:
1. Ensure that the TankMaster WinSetup program is up and running.
2. In the WinSetup workspace, click the right mouse button on the Rosemount 5900
device icon and select the Proof Test option:
Note
The simulated reference reflector must be configured
(1)
for the Guided Proof Test function
to be enabled.
3. Select the Simulated Reference Reflector check box (see “Maintenance” on
page 29).
4. Select the One-Point Level Verification check box.
5. Select one of the check boxes depending on the output option that is used:
Analog Output, or
Relays K1/K2 verification tests
1. See the Rosemount 5900 Proof Test Manual Supplement for more information.