16. When the One-point Verification test is finished, the proof test procedure
continues with the appropriate Output Verification test (as selected in step 5).
17. Depending on the selected option in step 5 (Proof Test window) see:
“Relay output verification” on page 28 or
“Analog output verification” on page 28.
18. When the complete proof test is finished, click the Next button to open the Proof
Test Summary window. You will have to sign this form in order to finish the test. A
report in PDF format will be created automatically. It may be printed and saved in
the desired network location. The report will be available from the Proof Test History
window
(1)
as well.
1. See the Rosemount 5900 Proof Test Manual Supplement for more information.