March 2018
10
Quick Start Guide
5.0 Verify configuration
Use the following block examples to do basic configuration to the transmitter.
For more advanced configurations see the Rosemount 3051S FOUNDATION
Fieldbus Reference Manual
.
Note
DeltaV
™
users should use DeltaV Explorer for the resource and transducer blocks and
Control Studio for the function blocks.
5.1 To configure the AI block
AI block configuration parameters
Use the Pressure, DP Flow, and DP Level examples for guides.
Parameters Enter data
Channel 1 = Pressure or 2 = Sensor temp
L_Type Direct, indirect, or square root
XD_Scale Scale and engineering units
Pa bar inH
2
O @ 68 °F cmH
2
O @ 4 °C
(1)
inHg @ 0 °C
kPa mbar mmH
2
O @ 68 °F ftH
2
O
(1)
mmHg @ 0 °C
MPa atm ftH
2
O @ 68 °F inH
2
O
(1)
mmH
2
O @ 4 °C
hPa
(1)
1. Only available with option IT6.
psf
(1)
inH
2
O @ 60 °F
(1)
g/cm
2
inH
2
O @ 4 °C
psi °C ftH
2
O @ 60 °F
(1)
kg/m2
(1)
mHg @ 0 °C
(1)
torr °F ftH
2
O @ 4 °C
(1)
kg/cm
2
cmHg @ 0 °C
(1)
mH
2
O @ 4 °C
(1)
Out_Scale Scale and engineering units