00809-0100-4835 Rev BE
Maintenance & Troubleshooting
SYMPTOM
CORRECTIVE ACTION
Low Output
or No Output
Replace terminal block.
Transmitter
Electronics Failure
Determine faulty circuit board by trying spare electronics
assembly or terminal block assembly. Replace faulty assembly.
Sensor Module
NOTE: See Sensor Module Checkout section. The sensing
element is not field repairable and must be replaced if found to
be defective. See Disassembly Procedure for instructions on
disassembly. Check for obvious defects (i.e. punctured isolating
diaphragm, etc.) and contact Rosemount Nuclear.
Check the power supply output voltage at the transmitter.
Erratic
Output
and Process
Check for entrapped gas in liquid lines, or liquid in dry lines.
Transmitter
Electronics
• Check for intermittent shorts or open circuits.
• Make sure the pins on the jumper, the pins on the filters, and
the pins on the sensor module are clean
Transmitter
Electronic Failure
Determine faulty circuit board by trying spare electronics
assembly or terminal block assembly. Replace faulty assembly.
Check power supply output voltage.