113
Rosemount 8600 Vortex Flow Meter
Reference Manual
May 2019
Electronics Verification
00809-0100-4860, Rev BF
URV= 5833 SCFM LRV= 0 SCFM F
in
(Hz) = 200.00
= 8.36 mA
Therefore, an input frequency of 200 Hz represents a current output of 8.36 mA.
I
F Hz K C
20
LRV–
URV LRV–
----------------------------------------------------------------
16
4+=
I
200 10.797 0.011641 0–
5833 0–
----------------------------------------------------------------------------------
16
4+=