eMKW-TD0054-21- Page 77
outweigh the risks to the device and more importantly, the patient.
Calibration
In order for the KUDUwave to operate correctly it is vital that it is calibrated correctly. eMoyo has the
equipment and necessary expertise to calibrate your KUDUwave. Please contact eMoyo to schedule your
next calibration.
Daily Calibration Verification
It is advisable to perform a biological test at the start of each day that the KUDUwave will be used.
Annual Calibration
Annual calibration is required. eMoyo will calibrate and return your KUDUwave to you in accordance
with your maintenance contract.
Note: The KUDUwave software has been designed to automatically warn you when the minimum
calibration is due. Your device will be assessed annually to determine its serviceability before it is
calibrated. The KUDUwave users can expect a minimum service life of 5 years.
Ad Hoc Calibrations
Some states or countries require audiometers to be calibrated more than once a year, especially when
audiometers are used for mobile testing. Please make sure what the legislative requirements are for your
state or country. If you are ever in doubt whether the KUDUwave is in calibration it is recommended to
verify the calibration biologically. If you are still in doubt it is essential to do an electro- acoustic
calibration check.
Calibrations by Other Organisations
Due to the digital, robust and integrated design of the KUDUwave, chances of it going out of calibration
are very slim. If another organisation wants to calibrate a KUDUwave, they may perform a typical
calibration process to verify that the thresholds and frequencies of the unit are correct. After completing
the calibration routine the calibration organisation will not need to adjust any thresholds or frequencies.
A calibration certificate must be issued to show that the KUDUwave is within calibration limits. In the
unlikely event where the calibration shows limits outside acceptable limits, please contact eMoyo
immediately.
Storage and Shipping
Warning: When shipping the KUDUwave please use the robust shock absorbing carry case to reduce the
risk of damage to the device during transit.
Remember to seal the shipping container securely and to mark the container FRAGILE. Always store the
KUDUwave out of direct sunlight in a clean and dry environment within the temperature and humidity
limits detailed in the Technical Specifications.