eMKW-TD0054-21- Page 88
Tympanometry General
KUDUwave™ built-in transducer
Influence of ambient pressure and
temperature on calibration
The KUDUwave™ Pro TMP contains an environmental sensor which measures
atmospheric pressure, temperature and relative humidity. Conversion between
volume and admittance is handled automatically. The unit will ask for recalibration
with the calibration cavities if the environment changes significantly.
Use only the tympanometry probes (identified on the packaging)
The probe should be visually inspected at each use. If it is dirty it should be cleaned
using the cleaning kit provided. If it shows signs of damage it should be replaced
with a new probe.
Probe Signal
85 dB SPL (≈ 69 dB HL) ±3 dB in an IEC 60318-5 coupler
Typical variation with loading:
6 dB at 0.5 cm³, 0 dB at 2 cm³, -6 dB at 5 cm³
Pneumatic system
200 daPa/s
Positive-to-negative and negative-to-positive
-750 daPa and +550 daPa as measured in a 0.5 cm³ cavity
Automatic valve release at safety maximum limits
±10 daPa or ±10%, whichever is greater (in cavities from 0.5 cm³ to 5 cm³)
200 and 400 daPa/s: ±40 daPa/s
(in cavities from 0.5 cm³ to 5 cm³)
Admittance Measurement
cm
3
or acoustic mmho (1 acoustic mmho = 10
-8
m
3
/(Pa⋅s)
0.2 cm³ to 5 cm³ (measurement plane)
±5% or ±0.1 cm³ of the equivalent volume or ±0.1 acoustic mmho, whichever is
greater.
This is applicable for both static and dynamic modes of operation
Dependence on barometric
pressure
The KUDUwave™ Pro TMP contains an environmental sensor which
automatically compensates for the conversion between volume and admittance