Water resistance
Very Low risk level – no harm – the device is splash proof and will operate normally aer contact
with sweat, rain or splashes. The device is not fully waterproof so the user should not submerge
it in liquid or wear it in the shower. In the event that there is water ingress into the case the User
will not be harmed, but the device may stop working.
Biocompability
Very Low risk level – no harm - the E4 is made of a well known biocompable material and the
manufacturer is qualied, ensuring that no contaminaon occurs during the manufacturing
process. Materials are outlined in Technical Specicaons secon.
Electromagnec compability
Very Low risk level – no harm – design validaon and vericaon has been conducted and has
demonstrated that the E4 is safe for use with other electronic devices. It is compliant with the
following standards:
• IEC 60601-1-2:2007 (third edion)
• FCCC FR47Part15bedIC (Industry Canada)
• ETSIEN301489-1e3V
There should be no precauons to be taken for any interference with the magnec elds, or to
the environmental condions of use.
Use Condion
Ambient - E4 can be used in domiciliary sengs, indoor or outdoor; it’s resistant to water but
cannot be immerged in water, can be lost by subject during use, should be worn ght to the
wrist.
Light condion - 100-500lx, Visual distance: 20-40 cm, Visual angle +-90°
Environmental condion - Temperature -10°C +40°C, Relave humidity 20% - 95%, Air pressure –
500hPa to 1200hPa, Acousc level - not relevant
Use frequency - It is extremely variable depending on subject/praconer prescripon:
Repeve use (once per day) or prolonged use (few minutes to 40hrs)