EasyManua.ls Logo

Empatica E4 - 8. Taking Care of Your Device; Cleaning and Disinfection; Types of Electrodes; Replacing the Electrodes

Empatica E4
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
empaca
©
2015
8. TAKING CARE OF YOUR DEVICE
Cleaning and disinfecon
Before switching between Users, and it is recommended that every week, carefully clean all parts
of the device that come in contact with the skin. The disinfecon processes should include one or
more of the following methods:
Exposure to UV lamp.
Cleansing with coon soaked with a small quanty of ethyl or isopropyl alcohol.
Cleansing with a soluon containing Alkyl C12-16 Dimethylbenzyl Ammonium Chloride.
Do not use cleaning agents other than those described above without explicit wrien instrucons
from Empaca support.
Types of electrodes
The E4 comes with two types of user-replaceable 8mm diameter electrodes: stainless steel and
silver (Ag). Please note that the Stainless steel electrodes are darker than the Ag ones.
Extensive tesng has demonstrated that the stainless steel electrodes are performing as well as
classic Ag electrodes when used with the E4. Stainless steel electrodes are hypoallergenic and
should last for the life of the device. The Ag electrodes should last for months under connuous
use but will oxidize, picking up a grey tarnish over me. The EDA electrodes should not be replaced
frequently as screwing and unscrewing repeatedly may degrade the performance of the sensor.
Reserve electrode replacement for end-of-life replacement of Ag electrodes or for electrode
comparison studies. It is not necessary to change electrodes between Users. The disinfecon
process can be done when the electrodes are aached to the E4.
Replacing the electrodes
When necessary the E4 electrodes should be changed by hand. Rotate them counter-clockwise to
loosen and clockwise to ghten. Make sure the electrodes are ghtened securely aer replacement
to ensure maximum data quality. Secure the electrodes by hand, not using a tool, to avoid damaging
the mechanical connecon to the E4 band.
If the electrode cannot be removed then the screw-post may be damaged. Report any problems to
support@empaca.com for further instrucons.
The Empaca E4 has a 2-year expected lifeme, considering the correct condions of use.