30 Continuum User’s Manual
TENS Custom — Summary
This treatment option is a sequenced, 2-channel Modulated Amplitude
(MOD) or a Simple Modulated Pulse (SMP) program for the management of
acute or chronic pain. All parameters have default settings and are adjustable
by the clinician within the range described in the table below.
Adjustable Parameters Range Default Setting
Treatment Time 5-60 min or Continuous 30 min
Pulse Rate 2 Hz-150 Hz 100 Hz
Cycling Time 3-20 sec 12 sec
Span (MOD Only) 20% - 95% 40%
Mode SMP, MOD SMP
Preset Parameters Settings
Pulse Width 300 µs
Pulsed DC Current Regimens
Pulsed DC Current (Edema) Acute — Summary
This treatment option is a sequenced, positively DC unbalanced (+) 2-channel
program for edema reduction and increased circulation. All parameters are
preset with the exception of Treatment time, which can be adjusted by the
clinician.
Adjustable Parameters Range Default Setting
Treatment Time 5-60 min or Continuous 30 min
Pulsed DC Current (Edema) Chronic — Summary
This treatment option is a sequenced, positively DC unbalanced (-) 2-channel
program for edema reduction and increased circulation. All parameters are
preset with the exception of Treatment time, which can be adjusted by the
clinician.
Adjustable Parameters Range Default Setting
Treatment Time 5-60 min or Continuous 30 min
Product Information (Clinician)