Continuum User’s Manual 31
Product Information (Clinician)
NMES Regimens
NMES large muscle:
•Amplitude at 1000 ohm is 100 V
peak
(DC component 2.2%)
•Amplitude at 500 ohm is 50 V
peak
(DC component 3.4%)
NMES small muscle:
•Amplitude at 1000 ohm is 100 V
peak
(DC component 2.1%)
•Amplitude at 500 ohm is 50 V
peak
(DC component 4.3%)
TENS Regimens
TENS SMP:
•Amplitude at 1000 ohm is 100 V
peak
(DC component 2%)
•Amplitude at 500 ohm is 50 V
peak
(DC component 3.4%)
TENS MOD:
•Amplitude at 1000 ohm is 100 V
peak
(DC component 2.3%)
•Amplitude at 500 ohm is 50 V
peak
(DC component 3.4%)
Pulsed DC Current Regimens
Acute Bi-Phasic:
•Amplitude at 1000 ohm is 100 V
peak
(DC component 1.6%)
•Amplitude at 500 ohm is 50 V
peak
(DC component 2.7%)
Acute Tri-Phasic:
•Amplitude at 1000 ohm is 100 V
peak
(DC component 2.3%)
•Amplitude at 500 ohm is 50 V
peak
(DC component 3.1%)
Chronic Bi-Phasic:
•Amplitude at 1000 ohm is 100 V
peak
(DC component 2%)
•Amplitude at 500 ohm is 50 V
peak
(DC component 3.7%)