Continuum User’s Manual 27
Small Muscle Custom / Remote — Summary
This option offers similar stimulation as Small Muscle Spasm and Trigger-
Point if no external Hand Switch is detected.
Adjustable Parameters Range Default Setting
Treatment Time
5-60 min or Continuous 30 min
O Time
0-60 sec 30 sec
Pulse Rate
2 Hz-150 Hz 35 Hz
Pulse Width
48 µs-400 µs 300 µs
Waveform Type
SYM, ASY ASY
Cycling Type
SIM, 1 CH, ALT, LAG SIM
Lag Delay
0-5 sec 0 sec
CH 1 Ramp+
0-5 sec 2 sec
CH 1 On Time
0-60 sec 10 sec
CH 1 Ramp–
0-5 sec 2 sec
CH 2 Ramp+
0-5 sec 2 sec
CH 2 On Time
0-60 sec 10 sec
CH 2 Ramp–
0-5 sec 2 sec
If used with an external Switch, continuous stimulation is available for gait
training protocol. The intensity of each channel is independently adjustable.
Waveform, pulse width, and pulse rate parameters have default settings
and are the adjustable parameters (within range). Depending on the cycle
selected, all parameters may not be adjustable.
TENS Regimens
TENS Knee — Summary
This treatment option is a 2-channel Modulated Amplitude program for the
management of acute or chronic pain control in the knee. All parameters are
preset with the exception of Treatment Time and the Pulse Rate, which can
be adjusted by the clinician.
Adjustable Parameters Range Default
Treatment Time 5-60 min or Continuous 30 min
Pulse Rate 2 Hz-150 Hz 100 Hz
Preset Parameters Settings
Cycling Time 12 sec
Attenuated Span 40%
Pulse Width 300 µs
Mode MOD
Product Information (Clinician)