42 Continuum User’s Manual
Timer
The timer is adjustable 5 to 60 minutes or unlimited.
Remote Control Hand Switch
Active regimen starts with closed contact.
Low Voltage Indication
•Indicator Threshold: 2.07 Volts (typical)
•Shutdown Voltage: 1.77 Volts (typical) 1.5 Volt Minimum
*These voltages may be tested under NO load condition.
Physical Characteristics
•Size: 1.26” (3.2cm) x 3.3” (8.4cm) c 4.5” (11.4)
•Weight (with batteries): 226.8 grams
Environmental Conditions
•Operational Temperature: 10° C to 40° C
•Humidity (Maximum): 10% - 90% RH
•Transport & Storage: Store in Dry, Cool Place
40° C to 70° C
•Pressure: 50-106 kPA
EN 60601-1 Classification
Type BF Applied Part
•Internally powered only
•Ordinary protection ingress of liquids
•Continuous operation
•Not suitable for use in the presence of a
flammable anesthetic mixture with air or
with oxygen or nitrous oxide.
Product Information (Clinician)