38 Continuum User’s Manual
Battery
Two AA rechargeable NiMH Batteries (Part Number: 200034-001)
Electrodes
Any Empi Electrode with a minimum active area of 16 cm
2
may be used with
this device. This includes Empi square (2” x 2”) StimCare electrodes.
DANGER: Use of an electrode with an area less than 16 cm
2
can cause
burns when the unit is used at higher intensities. Consult your
clinician prior to using any electrode less than 16 cm
2
.
Lead Wires
Any Empi TENS/NMES Lead Wire (100 cm – 40”) with the touch proof device
connection (part number 193057-100).
Remote Switches
•Use either: Empi Foot Switch - Part Number: 198887-001
•Empi Hand Switch - Part Number: 57202241
•Bifurcated Lead wire (for Edema) - Part Number - 700211-001
•Dispersive Pad (for Edema) - Part Number - 199501-001
Accessories
Standard Measurement Conditions
•Temperature - 25°C +/-5°C
•Load -1 kohm
•Power Supply - 3.0V DC +/- 10%
Typical Waveforms
The following are theoretical standard measurement output voltage across
purely resistive loads at maximum intensity setting. Pulse Width and Vpp
measured as shown across 500 ohm and 1k ohm loads. Your output may vary
depending on parameter settings.
Technical Data