Continuum User’s Manual 49
Troubleshooting
To self-test for any of the above, perform the following steps:
1. Place new batteries in the device.
2. Verify the device is off.
3. Insert one new lead wire into two new electrodes.
4. Place the new electrodes on your forearm as shown in the figure
below.
5. Insert the lead wire in Channel 1.
6. Turn your device on.
7. Select a TENS knee program. This is a continuous treatment program.
8. Slowly increase the amplitude (intensity) until you can feel it. If you
do not get any sensation, lower the amplitude (intensity) to zero
and rotate the lead wire 90 degrees. Slowly increase the amplitude
(intensity).
9. If there is no sensation, call the repair department.
10. If sensation is felt - even if weak, = device is working properly. You may
need to reposition the electrodes or contact your clinician.
11. Repeat Steps 1 through 10 for Channel 2.