6
02 03 04 AUTOMATIC CLEAN, DISINFECT AND DRY WITH WASHER DISINFECTOR
The washer-disinfector must comply with ISO 15883, must have suitable baskets to hold small products and must have
rinsing connections with diameter of approximately 16mm for the attachment to product lumen.
Correctly place the product into a suitable rack, connect all lumens to the rinsing connectors and start the automated cleaning.
Instructions of the manufacturer of the washer disinfector must also be observed.
Carefully follow the instructions provided by the disinfection solution manufacturer.
The following automated process
3
can be used to achieve an A0 Level of 3000:
2 min
Pre-clean with cold tap water
4
.
Drain
5 min
Clean at 55°C with tap water and 0.5% neodisher MediClean Dental (Dr Weigert) solution.
Drain
3 min
Rinse and neutralize with cold demineralized water
5
.
Drain
2 min
Rinse with cold demineralized water
5
.
Drain
3 min
Perform thermal disinfection (nal rinse) with deionized water at 93°C minimum.
Drain
20 min (at least)
Dry at 100°C.
02 MANUAL CLEAN +
03 DISINFECTION + 04 DRY
The following validated EMS process can be used with and without ultrasound.
DO NOT use any ultrasound bath cleaning procedure with the handpieces: it may destroy the products. It is ONLY intended
to be used with EMS PIEZON
®
Instruments and le holders. In case of ultrasound bath cleaning, place products in a metal sieve
and sonicate 10 minutes.
Validated
product
Process
EMS products
to clean/
disinfect
Process
0.5%
neodisher
MediClean
(Dr. Weigert)
without
ultrasound
bath
all EMS
products
Brush the devices in a cleaning solution of 0.5% neodisher in deionized water at 40°C
with a suitable soft bristled brush until all visible residue has been removed.
15 s
For products with lumens, flush all lumens with a spray gun (water jet
gun, with a static water pressure of 2 bar) with cold tap water.
15 min
(10 min with
ultrasound)
Place the product in 0.5% solution of neodisher with deionized water at
40°C. Make certain that all lumens are lled with cleaning solution (use
a syringe if needed).
15 s
Rinse all lumens by flushing with a spray gun (water jet gun, with a static
water pressure of 2 bar) with cold deionized water
5
.
10 s
Rinse the whole product under cold running deionized water.
At room temperature (15-25°C), use an air pistol (compressed air, max. 3 bars) to
fully dry the lumen and the whole product until no more residues of water are present
(visible or detectable).
0.5%
neodisher
MediClean
(Dr. Weigert)
with
ultrasound
bath
6
EMS PIEZON
®
Instruments
and le
holders
Then disinfect:
ASP CIDEX
®
OPA solution must be used undiluted, within its useful and shelf life, and in compliance with the manufacturer’s
warnings and instructions for use.
5 min
Immerse the product completely into CIDEX
®
OPA solution at minimum 20°C. Make certain that all lumens are lled
with disinfectant solution (use a syringe if needed).
1 min
(each rinse)
The ASP CIDEX
®
OPA disinfectant requires a total of three rinses : keep the product totally immersed, and use a large
volume
7
of fresh water.
Do not reuse the water for rinsing or any other purpose. Residues of disinfectant may cause serious side effects.
And dry
At room temperature (15-25°C), use an air pistol (compressed air, max. 3 bars) to fully dry the lumen and the whole product until
no more residues of water are present (visible or detectable).
05 INSPECT
If stains are still visible on the product after cleaning/disinfection, the entire cleaning/disinfection procedure must be
repeated. Products with visible damage, chip/flake loss, corrosion or deformation must be disposed of (no further use is
allowed). Check also the integrity of O-rings and gaskets and replace if damaged or deformed.
Verify the product to be fully dry. In case of detection of residues of water, remove these using an air pistol (clean compressed
air). Fully dry the lumen and the whole product until no more residues of water are present (visible or detectable).
3 For example: Miele Professional G 7836 CD with Miele Rack E429
4 cold tap water = 16°C +/- 2°C
5 cold demineralized water = 20°C +/- 2°C
6 For example: Bandelin, Sonorex 1028K, 35kHz
7 e.g. 2 gallons
OR
EITHER

Other manuals for EMS AIRFLOW EL-185

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW EL-185 and is the answer not in the manual?

EMS AIRFLOW EL-185 Specifications

General IconGeneral
BrandEMS
ModelAIRFLOW EL-185
CategoryDental equipment
LanguageEnglish

Related product manuals