E2 MicroProbe¥ System Turn-on
1. Turn the mains power switch “On”. A system Self-Test begins.
2. The audible Pre-Laser Firing (PLF) Warning Alarm will remain ON for
approximately 10 seconds.
3. After the PLF Warning Alarm ceases, the system will go into the Standby
mode.
4. The STANDBY light on the display panel illuminates after the PLF Warning
Alarm and the COOLDOWN light go off (if it illuminates).
The E2 MicroProbe¥ is now ready for Setting of the Treatment Beam.
E2 MicroProbe¥ Setting of the Treatment Beam
To set the treatment beam:
1. Insert the key in the Key Switch and turn to the "On" position.
2. Depress the Laser Power push button until the desired laser power appears
in the Laser Power display. The lowest practical setting should always be
used.
3. Depress the Laser Duration push button until the desired exposure time
appears in the Laser Duration display.
4. Press the Enable push button and the green LED status indicator will light.
The 810 NM laser will not fire until the Foot Pedal is used.
SAFETY PRECAUTION
The LASER DURATION and the LASER POWER must be set to get an
aiming beam.
Before Firing the Laser
1. Adjust the Light Source push buttons to obtain a satisfactory video image. If
the Light Source is set too high, the aiming beam will appear diminished.
2. Bring the target video image into clear focus by rotating the star wheel on the
Video Adapter while viewing the target through the endoscope.
3. Place the tip of the endoscope near a white paper. Press the Aiming Beam
Intensity push button up and verify that the red aiming beam intensity
increases on the paper and the monitor. An audible modulating tone
indicates an aiming beam firing.