EMC TECHNICAL INFORMATION
The E4 Endoscopy System needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided
in the following information.
Guidance and manufacturer's declaration - electromagnetic emissions
The E2 Microprobe Laser and Endoscopy System is intended for use in the electromagnetic
environment specified below. The customer or the user of the E2 Microprobe Laser and Endoscopy
System should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The E2 Microprobe Laser and Endoscopy System uses RF energy
only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
The E2 Microprobe Laser and Endoscopy System is suitable for use
in all establishments other than domestic ,and may be used in
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes, provided the following warning is heeded:
Warning: This equipment is intended for use by healthcare
professionals only. This equipment may cause radio interference or
may disrupt the operation of nearby equipment. It mat be necessary
to take mitigation measures, such as re-orienting or relocating the
E2 Microprobe Laser and Endoscopy System or shielding the
location.