TECHNICAL SPECIFICATIONS
Light Source
175 Watt Xenon Light
x Visible Output 2,200 Lumens
x Radiant Output 25 Watts
300 Watt Xenon Light
x Visible Output 5,000 Lumens
x Radiant Output 50 Watts
Expected lifetime of the light source is 1,000 hours until the light reaches half
of its initial output. (This could be equivalent to up to 6 years of usage.
Calculations are based on 3.25 hours of use per week.)
The failure of the light source is expected to be the deciding factor of the service
life of the equipment.
CAUTION: The heat generated by the light source varies and can be hot causing
the light connector, on the endoscope, to become hot. To prevent injury to the
operator the light intensity should be turned down so the connector can cool
before removing it from the system.
Treatment Laser
•
Semiconductor GaAlAs Laser Diode, 810 r 25 Nanometers,
• Waveform can be continuous or intermittent
• 1.2 Watts Output
• Class IV Laser
Nominal Ocular Hazard Distance - 0.6 Meters
Aiming Beam Laser
•
Visible Laser Diode, 640 r 30 Nanometers,
• 1.5 Milliwatt Output
• Class IIIR Laser
Operational Mode
•
Continuous Wave
Laser Source Power
•
Less than 5 watts
Endoscope
•
Beam Divergence - 40q
• Laser Delivery System - 200 Pm fiberoptic terminating in a 0.89 mm diameter distal
end SS tube with a laser fiber, video fiber and light fiber inside.
• Fiberoptic Connector - SMA 905 with the Endo Optiks Optical Safety Interlock to
prevent firing the laser without the Endoscope connected.
• Nominal Ocular Hazard Distance - 0.6 Meters
Console Cooling
•
Forced air