Cubemass DCI
Endress+Hauser 5
Input
Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
• Volume flow (calculated using mass flow and density)
• Fluid density (proportional to the resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)
Measuring range Measuring ranges for liquids
Operable flow range 1:100
Input signal Status input (auxiliary input)
U = 3 to 30 V DC, R
i
= 5 kΩ, galvanically isolated.
Switching level: 3 to 30 V DC, polarity-independent.
Configurable for: totalizer reset, positive zero return, error message reset, start zero point adjustment.
DN Range for full scale values (liquids) g
min(F)
to g
max(F)
[mm] [inch] [kg/h] [lb/min]
1 1/24" 0 to 20 0 to 0.75
2 1/12" 0 to 100 0 to 3.7
4 1/8" 0 to 450 0 to 16.5
6 ¼" 0 to 1000 0 to 37