Cubemass DCI
Endress+Hauser 5
Input
Measured variable Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
Volume flow (calculated using mass flow and density)
Fluid density (proportional to the resonance frequency of the measuring tube)
Fluid temperature (measured with temperature sensors)
Measuring range Measuring ranges for liquids
Operable flow range 1:100
Input signal Status input (auxiliary input)
U = 3 to 30 V DC, R
i
= 5 kΩ, galvanically isolated.
Switching level: 3 to 30 V DC, polarity-independent.
Configurable for: totalizer reset, positive zero return, error message reset, start zero point adjustment.
DN Range for full scale values (liquids) g
min(F)
to g
max(F)
[mm] [inch] [kg/h] [lb/min]
1 1/24" 0 to 20 0 to 0.75
2 1/12" 0 to 100 0 to 3.7
4 1/8" 0 to 450 0 to 16.5
6 ¼" 0 to 1000 0 to 37