Endress+Hauser Deltapilot M FMB51 - 6.5.7 Calibration With Partially-Filled Tank

Endress+Hauser Deltapilot M FMB51
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Cerabar M, Deltabar M, Deltapilot M / 4...20 mA HART
66 Endress+Hauser
6.5.7 Calibration with partially-filled tank (wet calibration)
Example:
In this example a wet calibration is shown when it is not possible to empty the vessel and then fill
it up to 100%. Here a 20% filling is used as “Empty” and a “25%” filling is used as “Full” calibration
point. The calibration is then extended to 0% ... 100% and LRV / URV are adjusted accordingly.
Prerequisite:
The default value in the level mode for calibration mode is “Wet”.
However, it can be changed via: Setup Extended Setup Level Calibration mode
!
Note!
It is also possible to use different liquids (e.g. water) for the adjustment. In this case you have to
enter the different densities at following menu path:
•Setup Ext. Setup Level Adjust density (034) (e.g. 1.0 kg/l for water)
•Setup Ext. Setup Level Process density (035) (e.g. 0.8 kg/l for oil)
Description
1 Select the "Level" measuring mode via the "Measuring
mode (005)" parameter.
Menu path: Setup Measuring mode (005)
P01-Mxxxxxxx-19-xx-xx-xx-007
Fig. 35: Calibration with partially-filled tank
1 See Table, Step 2
2 See Table, Step 3
2 Set value for "Empty calib." with acting pressure for
Level e.g. 20%.
Menu path: Setup Extended Setup Level Empty
calibration
3 Set value for "Full calib." with acting pressure for Level
e.g. 25%.
Menu path: Setup Extended Setup Level Full
calibration
4 The values for full and empty pressure are measured
automatically at adjustment.
As the transmitter automatically sets the pressure values
that suit to empty and full calibration to min and max
pressure that cause the output current, it is necessary to
set the right upper range value (URV) and lower range
value (LRV).
20%
20%
1
25%
2

Table of Contents