EasyManua.ls Logo

Endress+Hauser Flowphant T DTT31 - About This Document; Document Function and Symbols

Endress+Hauser Flowphant T DTT31
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DTT31, DTT35
4 Endress+Hauser
Multi Channel Recorder Ecograph T
144 x 144 mm (5.67 x 5.67 in) panel mounted housing for electronic acquisition, display, recording,
analysis, remote transmission and archiving of analog and digitalinput signals. Data recording system with
CompactFlash card and multi-coloured LCD display (120 mm / 4.7 in screen size). Configuration and
visualisation via interface (USB, Ethernet, RS232/485) and ReadWin
®
2000 PC operating software.
Data logger Minilog B
Battery powered measured value collector with 2 input channels for storing analog and digital values.
Internal memory 128 kB for max. 84000 measured values. Configuration and visualisation via RS232
interface and ReadWin
®
2000 PC operating software. Optionally with telealarm function.
Input
Measured variable Flow velocity of liquid media (calorimetric measuring principle)
Temperature (RTD), optional for two switch outputs or additional analog output
Measuring range Flow: 0.03 m/s to 3 m/s (0.1 to 9.84 ft/s), as relative value between 0 and 100%; maximum display
resolution: 1%
Temperature: -20 °C to +85 °C (-4 to +185 °F); display resolution: 1 °C (1 °F)
Output
Output signal DC voltage version: (short-circuit proof version)
1x PNP switch output (flow) or
2x PNP switch outputs (flow or temperature, adjustable) or
1x PNP switch output and 1x 4 to20 mA output, active (flow or temperature, adjustable)
!
Note!
The analog output reads out the measured flow rate as relative value in percentage of the adjusted measuring
range.
Signal on alarm Signal on alarm as per NAMUR NE43
Underranging: Linear drop to 3.8 mA
Overranging: Linear rise to 20.5 mA
Sensor break; sensor short-circuit:
3.6 mA or 21.0 mA (at settings 21.0 mA, 21.7 mA output is guaranteed)
Switch outputs: at safety condition (switch open)
Load Max. (V
power supply
- 6.5 V) / 0.022 A (current output)
Range of adjustment Switch output
Switch point (SP) and switchback point (RSP) in increments of 1% with min. hysteresis of 5%
Damping: freely adjustable: 0 = off (no damping) or 10 to 40 s in increments of 1 second
Unit: %, optional °C, °F (with two outputs and temperature monitoring)
Switching capacity DC voltage version:
Switch status ON: I
a
250 mA, switch status OFF: I
a
1 mA
Switching cycles: > 10,000,000
Voltage drop PNP: 2 V
Overload protection
Automatic testing of switching current; output is switched off in case of overcurrent, the switching current
is tested again every 0.5 s; max. capacitance load: 14 μF for max. supply voltage (without resistive load)
Periodic disconnection from a protective circuit in event of overcurrent (f = 2 Hz) and indication of
’Warning’

Other manuals for Endress+Hauser Flowphant T DTT31