EasyManua.ls Logo

Endress+Hauser Flowphant T DTT31 - Mounting and Installation; Mounting Requirements and General Installation

Endress+Hauser Flowphant T DTT31
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DTT31, DTT35
Endress+Hauser 9
Operating conditions: Environment
Ambient temperature range –40 up to +85 °C (-40 to +185 °F)
Storage temperature –40 up to +85 °C (-40 to +185 °F)
Climate class 4K4H as per DIN EN 60721-3-4
Degree of protection IP 65 (complete housing)
Shock resistance 50 g as per DIN IEC 68-2-27 (11 ms)
Vibration resistance 20 g as per DIN IEC 68-2-6 (10-2000Hz)
4 g as per German Lloyd GL Guidelines
Electromagnetic compatibility Interference emission as per IEC 61326 Series, class B electrical equipment
Interference immunity as per IEC 61326 Series, appendix A (industrial use) and NAMUR Recommendation
NE 21
EMC influence: 0.5 %
Operating conditions: Process
Process temperature limits -20 to +85 °C (-4 to +185 °F)
!
Note!
The sensor can be exposed up to 130 °C (266 °F) without damage; monitoring switches off automatically at
T 85 °C (185 °F) and starts again at T 85 °C (185 °F).
Process pressure limits Maximum permitted process pressure p
max
10 MPa = 100 bar (1450 psi)
"
Caution!
The maximum process pressure for the conical metal-metal process connection (MB option) for the DTT35 is
1.6 MPa = 16 bar (232 psi)!
Process flow limit Liquids: 0 up to 3.0 m/s (0 up to 9.84 ft/s)
Operating range Liquids: 0.03 up to 3.0 m/s (0.1 up to 9.84 ft/s)