KA651F/00/a2/09.10
Endress+Hauser 3
Exchange of on FMG60NaI-crystal sensor assembly
+-
+-+-
IN CASCADE OUT
DISPLAY
SIM WP
Pt100
OFF
ON
3
4
56
7
8
9
10
11
12
13
L1
L+
N
L-
+
-
1
2
(3)
(4)
15
Z
AB
14
Z
,
D
C
E
Terminal
compartment A
Holding time>2s
6. Testing
Insert jumper 5-9
+-
+-+-
IN CASCADE OUT
DISPLAY
SIM WP
Pt100
OFF
ON
3
4
56
7
8
9
10
11
12
13
L1
L+
N
L-
+
-
1
2
(3)
(4)
15
Z
AB
D
14
Z
C
Holding time>2s
Holding time > 2 s
4. Testing
Remove jumper
2-(3)
5. Testing
Remove jumper
4-5, 3-9
E
,
Holding time>2s
3. Testing
+-
+-+-
IN CASCADE OUT
DISPLAY
SIM WP
Pt100
OFF
ON
3
4
56
7
8
9
10
11
12
13
L1
L+
N
L-
+
-
1
2
(3)
(4)
15
Z
AB
D
14
Z
E
Test procedures for instruments with PROFIBUS PA or FIELDBUS Foundation electronics
+-
+-+-
IN CASCADE OUT
DISPLAY
SIM WP
Pt100
OFF
ON
3
4
56
7
8
9
10
11
12
13
L1
L+
N
L-
+
-
1
2
(3)
(4)
15
Z
AB
D
14
Z
C
,
1. Testing2. Testing
E
C
Note: To perform the following testing insert the jumper 5 - 9.
Note: To perform the following testings remove the jumpers 2 -(3),4-5,3-9.
,
Note: Never perform these testings on an instrument with HART electronics, the current output will be damaged!
Terminal
compartment B
Terminal
compartment A
Terminal
compartment B
Terminal
compartment A
Terminal
compartment B
Terminal
compartment A
Terminal
compartment B
600 Veff
848 V=
>340kW
1.5 kVeff
2.12 kV=
>850kW
600 Veff
848 V=
>340kW
2.3 kVeff
3.25 kV=
1.3 MW
2.3 kVeff
3.25 kV=
1.3 MW
600 Veff
848 V=
>340kW
Holding time > 60 s
Holding time>2s