TMT162 Operation
Endress+Hauser 15
5.2 Local operation
5.2.1 Setup of the hardware
Fig. 10: Hardware setup using jumpers J1, J2 and J3
"
Caution!
Protect the terminals from electrostatic discharge. Failure to observe this may result in destruction
of parts of the electronics.
Jumpers J1, J2 and J3 for the hardware setup can be found in the electronic chamber. In order to
set the jumper, open the electronics chamber screwed cover (opposite the connection chamber
cover) and if necessary pull off the display.
Setup or configuration hardware lock using jumper J1
The hardware setup/configuration lock has priority over the software setup.
TRANSMITTER SECURITY
ON Setup/configuration locked
OFF Setup/configuration unlocked