TMT162 Installation
Endress+Hauser 9
3.3.2 Remote installation
a0003586-ae
Fig. 4: Installation of the field transmitter using the mounting kit, see chapter 'Accessories' (dimensions in inches;
mm)
Pos. A, B Mounting with combined wall/pipe mounting kit
Pos. C Mounting with pipe mounting kit 2"/V4A
3.4 Installation check
After installing the device, always run the following final checks:
Device condition and specification Hint
Is the device visibly damaged (visual check)? -
Does the device comply with the measurement point specifications, such as
ambient temperature, measurement range, etc.?
See chapter 10 "Technical data"
Make sure that the transmitter covers are tight. Both transmitter covers must be
fully engaged to meet explosion-proof requirements.
-