1 Safety notes TMT182
10 Endress+Hauser
3.2 Installation
3.2.1 Typical European installation
Installation For installation, proceed as follows:
Installing in a sensor connection head to DIN 43 729 Form B (see fig. 4, left side)
Feed the sensor inset cables (Pos. 5) through the central hole in the head transmitter (Pos. 4).
Place the installation springs (Pos. 3) onto the screws (Pos. 2).
Feed the installation screws (Pos. 2) through the holes in the head transmitter and the holes in
the sensor inset (Pos. 5). Fix both screws using the E-rings (Pos. 6).
Position the head transmitter in the connection head in such a way so that the current output
terminals (terminal 1 and 2) are towards the cable entry gland (Pos. 7).
Fix the head transmitter (Pos. 4) and sensor inset (Pos. 5) into the connection head.
Installation in a field housing TAF10 (see fig. 4, right side)
Feed the installation screws (Pos. 2) with installation springs (Pos. 3) through the holes in the
head transmitter (Pos. 4). Fix these using the E-rings (Pos. 5).
Screw the head transmitter into the field housing using a screwdriver.
"
Caution!
In order to avoid damaging the head transmitter, do not over-tighten the installation screws.
ƒig. 4: Installation of transmitter into a sensor connection head (left side) and a field housing TAF10 (right side).