EasyManuals Logo

Endress+Hauser iTEMP TMT162 Technical Information

Endress+Hauser iTEMP TMT162
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
TI086R/24/ae/03.09
Technical Information
iTEMP
®
TMT162
Temperature field transmitter
with HART
®
, FOUNDATION™ Fieldbus or
PROFIBUS
®
PA protocol
Application
Input: RTD, TC, Ω, mV
•Output:
HART protocol for converting various input signals
to a scalable 4 to 20 mA analog output signal.
Operating the transmitter using handheld terminals
Field Xpert SFX100, DXR275/375 or remotely via
PC
FOUNDATION Fieldbus ITK 4.61
PROFIBUS PA
Optional: stainless steel housing for hygienic or Ex d
application
Features and benefits
High reliability in harsh industrial environments due to
dual compartment housing and compact, fully potted
electronics
Backlit display with large measured value, bargraph
and fault condition indication for ease of reading
Ability to display measured values of up to 4 other
devices (FOUNDATION Fieldbus)
Dual sensor input capability, e.g. 2 Pt100 3-wire or 1
Pt100 4-wire and thermocouple
Reliable operation due to sensor monitoring:
breakdown information, sensor backup, drift alarm
and corrosion detection
High measuring point accuracy due to sensor
transmitter matching: linearization with Callendar van
Dusen or polynomial equation
Operation voltage monitoring for high measurement
reliability (HART)
Mathematic functions for differential and average
temperature add flexibility to the measurement
FISCO/FNICO compliant according to IEC 60079-27
International approvals guarantee safe operation in
hazardous area: FM, CSA (IS, NI, XP and DIP) and
ATEX (Ex ia, Ex nA nL, Ex d and dust-Ex)
Galvanic isolation 2 kV (sensor input to the output)
Output simulation for a quick and easy loop check
1 0 â

Other manuals for Endress+Hauser iTEMP TMT162

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser iTEMP TMT162 and is the answer not in the manual?

Endress+Hauser iTEMP TMT162 Specifications

General IconGeneral
TypeTemperature Transmitter
CommunicationHART
Accuracy± 0.1 % of span
InputRTD
RTD TypePt100, Pt1000
Output4-20 mA
Measurement RangeConfigurable
Protection ClassIP66, IP67
Ambient Temperature-40 to +85 °C
ConfigurationVia HART communicator
ApprovalsATEX, IECEx, FM, CSA

Related product manuals