Operating options iTEMP TMT82
26 Endress+Hauser
ON
OFF
1
2
4
8
16
32
64
HW
SW
ADDR ACTIVE
SIM
WRITE LOCK
DISPL. 180°
1
2
3
A0014562
 17 Hardware settings via DIP switches
1: Connection to head transmitter
2: DIP switches (1 - 64, SW/HW, ADDR and SIM = simulation
mode) no function for this head transmitter
3: DIP switch (WRITE LOCK = write protection; DISPL. 180° =
switch, turn the display monitor 180°)
Procedure for setting the DIP switch:
1. Open the cover of the terminal head or field housing.
2. Remove the attached display from the head transmitter.
3. Configure the DIP switch on the rear of the display accordingly. In general: switch to ON
= function enabled, switch to OFF = function disabled.
4. Fit the display onto the head transmitter in the correct position. The head transmitter
accepts the settings within one second.
5. Secure the cover back onto the terminal head or field housing.
Switching write protection on/off
Write protection is switched on and off via a DIP switch on the rear of the optional attachable
display. When write protection is active, parameters cannot be modified. A lock symbol on the
display indicates that write protection is on. Write protection prevents any write access to the
parameters. Write protection remains active even when the display is removed. To disable the
write protection, the device must be restarted with the display attached and the DIP switch
deactivated (WRITE LOCK = OFF). Alternatively, the display can be removed and reattached
during operation to disable write protection.