Levelflex FMP56, FMP57
16 Endress+Hauser
4-wire, 4-20mA HART, active
"Power supply; Output"
1)
Terminal voltage
K: 4-wire 90-253VAC; 4-20mA HART 90 to 253 V
AC
(50 to 60 Hz)
L: 4-wire 10,4-48VDC; 4-20mA HART 10.4 to 48 V
DC
1) Feature 020 of the product structure
PROFIBUS PA
"Power supply; Output"
1)
Terminal voltage
G: 2-wire; PROFIBUS PA, switch output 9 to 32 V
DC
1) Feature 020 der Produkstruktur
Terminals
Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)
Cable entries
• Cable gland (not for Ex d):
–
Plastics M20x1,5 with cable Æ 5 to 10 mm (0.2 to 0.39 in): non-Ex, ATEX/IECEx/NEPSI Ex ia/ic/nA
–
Metal M20x1,5 with cable Æ 7 to 10 mm (0.28 to 0.39 in): dust-Ex, FM IS, CSA IS, CSA GP
• Thread for cable entry:
– ½" NPT
– G ½"
– M20 × 1.5
• Connector (only for non-Ex, Ex ic, Ex ia): M12 or 7/8"
Cable specification
•
For ambiente temperature T
U
³60 °C (140 °F): use cable for temperature T
U
+20 K.
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the plant.
Device plug connectors
For the versions with fieldbus plug connector (M12 or 7/8"), the signal line can be connected without
opening the housing.
Pin assignment of the M12 plug connector
A0011175
Pin Meaning
1 Ground
2 Signal +
3 Signal -
4 not connected
Pin assignment of the 7/8" plug connector
A0011176
Pin Meaning
1 Signal -
2 Signal +
3 not connected
4 Ground
Power consumption
min. 60 mW, max. 900 mW