Levelflex FMP56, FMP57
Endress+Hauser 19
Differing from this, the following measuring error is present in the vicinity of the lower probe end:
rodprobeandcoaxprobe
-80
-60
-40
-20
0
20
40
60
80
0 50 100 150 200 250 300
distancefromprobeend[mm]
sum of non-linearity, non-repeatability
and hysteresis [mm]
DC=2
DC>7
ropeprobe
-80
-60
-40
-20
0
20
40
60
80
0 50 100 150 200 250 300
distancefromprobeend[mm]
sum of non-linearity, non-repeatability
and hysteresis [mm]
DC>7
A0014154-EN
If for rope probes the DC value is less than 7, then measurement is not possible in the area of the straining
weight (0 to 250 mm from end of probe; lower blocking distance).
Differing from this, the following measuring error is present in the vicinity of the upper probe end
(rod/rope only):
3mm
0mm
-3mm
100mm 200mm
10mm
-10mm
20mm
-20mm
30mm
-30mm
40mm
-40mm
referencepointofmeasurement
maximummeasurederror
A0014155-EN
Resolution
digital: 1 mm
analog: 1 mA
Reaction time
The reaction time can be parametrized. The fastest possible reaction time is given by the measuring rate:
Probe length
Level measurement
<10 m (33 ft)
3.6 measurements/second
up to 40 m (131 ft)
³2.7 measurements/second