Liquicap M FTI51, FTI52 Installation
Endress+Hauser 25
Rope probes FTI52 (fully insulated)
!
Note!
The active probe length is always fully insulated (dimension L1).
Total length of probe from sealing surface: L = L1 + L3
All rope probes are prepared for tensioning in containers (tensioning weight with anchor hole)
Not suitable for agitator tanks, high-viscosity liquids and plastic tanks
Thickness of rope insulation 0.75 mm
Rope probe Rope probe with
clad Tri-Clamp
Rope probe
with inactive length
(uninsulated)
Rope probe with
fully insulated
inactive length
L00-FMI5xxxx-06-05-xx-xx-070
L00-FMI5xxxx-06-05-xx-xx-036
Total length (L) 420 to 10000 570 to 12000 570 to 11000
Active rope length (L1) 420 to 10000 420 to 10000 420 to 10000
Inactive length (L3) -- 150 to 2000 150 to 1000
Probe rope diameter 4 4 4
ø Anchor weight 22 22 22
ø Anchor hole 5 5 5
Tensile loading capacity (N) of probe rope at
20 °C
200
200 200
For aggressive liquids X -- X
For use in mounting nozzles -- X X
In the event of condensate on tank ceiling -- X X
For high-viscosity liquids -- -- --
X = recommended
PTFE
PFA
316L
120
L1
120
L1
L3
120
L1
L3
120
L1
Length tolerances L1, L3 up to 1 m: 0 to –10 mm 1 m to 3 m: 0 to –20 mm 3 m to 6 m: 0 to –30 mm 6 m to 12 m: 0 to –40 mm