Liquicap M FTI51, FTI52 Installation
Endress+Hauser 41
3.8.2 Wall mounting
Push the bracket onto the sleeve and screw it into place.
Mark the distance between holes on the wall, then drill the holes.
Screw the separate housing to the wall.
L00-FMI5xxxx-03-05-xx-xx-010
3.8.3 Pipe mounting
Push the bracket onto the sleeve and screw it into place.
Screw the separate housing to the pipe (max. 2").
L00-FMI5xxxx-03-05-xx-xx-011
3.9 Post-installation check
After installing the measuring device, carry out the following checks:
Is the device damaged (visual inspection)?
Does the device meet the specifications at the measuring point, including process temperature
and pressure, ambient temperature, measuring range etc.?
Is the process connection tightened with the correct torque?
Are the measuring point number and labeling correct (visual inspection)?
Is the measuring device adequately protected from precipitation and direct sunlight?