Liquiline CM442R/CM444R/CM448R Electrical connection
Endress+Hauser 25
5.2.4 Connecting the functional ground
You must always connect the terminal strip with PE from the central node in the cabinet.
Use the conductor with cable clamp that is included with the Memosens cable to connect the
functional earth to the terminal strip of the device.
GN/YE
 9 Functional earth connection
You must only ever connect one functional earth to each screw in the terminal strip.
Otherwise, shielding is not guaranteed.
5.2.5 Connecting sensors for non-hazardous area
Connection
Direct connection of the sensor cable to the terminal connector of the sensor module 2DS or of
the base module-L, -H or -E (→  10 ff.)