Parameters according to
IEC 61508
E+H Memosens pH SIL
measuring chain
Safety Function
1: pH limit monitoring
2: pH value measurement
3+4: safe calibration and adjustment
SIL
Hardware: 2, Software: 2
in homogenous redundancy: 2
HFT 0
Device Type B
Mode of Operation Low demand mode
SFF 93.8 %
MTTR (used for PFD calculation) 8 h
T
1
(Proof test interval)
Recommended: 1 / 1 / 1 year,
(sensor / cable / transmitter)
SD
688 FIT
SU
1623 FIT
DD
4473 FIT
DU
447 FIT
Total
*1
7238 FIT
PFD
avg
(for T
1
= 1 / 1 / 1 year)
*4
19.6 × 10
-4
PFH
4.5 × 10
-7
MTBF
*1
15 years
Diagnostic-Test-Interval
*2
< 60 min
Error Reaction Time
*3
< 10 seconds
DC
D
(Diagnostic Coverage Dangerous) 91 %