XA00182F-F Liquiphant M, Liquiphant S FTL50(H)/51(H)/51C, FTL70/71
24 Endress+Hauser
Non-sparking
Ex nA, Ex nC, Ex tc
In potentially explosive atmospheres: Do not disconnect electrical connections when energized.
Potential equalization
Integrate the device into the local potential equalization.
Explosion protection
with heat insulation
Device type FTL70, FTL71
While observing the "temperature derating", the device is suitable for process temperatures up to
300 °C.
When operating, ensure that you rule out contact between hot component surfaces and potentially
explosive atmospheres beyond the limits of the corresponding temperature class.
Suitable measures: e.g. thermal insulation at container and/or pipes.
The temperature of 85 °C specified at the reference point may not be exceeded.
To protect the electronics, observe the specified ambient temperature at the electronics housing.
3
4
5
2
1
T
a
T
p
A0025541
 6
T
a
Ambient temperature
T
p
Process temperature
1 Sensor
2 Temperature class, e.g. T6
3 Housing
4 Reference point: max. +85 °C
5 E.g. thermal insulation
Temperature tables Ambient temperature electronics
Temperature
class
Basic specification, Position 7
(Electronics; Output)
Ambient temperature T
a
(ambient)
T6 1, 2 –50 °C ≤ T
a
≤ +60 °C
T5-T2 –50 °C ≤ T
a
≤ +70 °C
T6 4 –50 °C ≤ T
a
≤ +40 °C
T5 –50 °C ≤ T
a
≤ +55 °C
T4-T2 –50 °C ≤ T
a
≤ +70 °C
T6-T2 5, 6, 7, 8 –50 °C ≤ T
a
≤ +70 °C
T6 A, D, 9 –50 °C ≤ T
a
≤ +60 °C
T5-T2 –50 °C ≤ T
a
≤ +70 °C