Commissioning Liquisys M CCM223/253
54 Endress+Hauser
Coding Field Range of
adjustment
(factory settings
in bold)
Display Info
O331 Select table
option
Read
Edit
A0025042-EN
O332 Enter number of
table value pairs
1
1 to 10
A0025043-EN
Enter the number of pairs of x
and y values (measured value
and current value) here.
O333 Select table
value pair
1
1 to no. elem.
Assign
A0025044-EN
The system runs through the
O333 to O335 function chain
as often as indicated in O332.
"Assign" appears as the last
step. The display goes to O336
after confirmation.
O334 Enter x value 0 ppb
0 to 20000 ppb
0.00 ppm
0 to 20 ppm
0.00 mg/l
0 to 20 (5) mg/l
pH 4.00
pH 4 to 9
0 mV
0 to 1500 mV
0 °C
0 to 50 °C
A0025046-EN
x value = measured value
specified by user.
O335 Enter y value 0.00 mA
0.00 to 20.00 mA
A0025048-EN
y value = user-defined current
value pertaining to O334.
Return to O333 until all values
are entered.
O336 Message as to
whether table
status is OK
Yes
No
A0025049-EN
Back to O3.
If status = no, correct the table
(all settings made up until now
remain intact) or go back to
the measuring mode (table is
deleted).