Micropilot FMR50/51/52/53/54/56/57 XA00677F-A
Endress+Hauser 19
Connection data Optional specification, ID Nx (Accessory Mounted) = NA
(Overvoltage protection (TRC [16], type OVP10 and TRC [17], type OVP20)
• When using the internal overvoltage protection: No changes to the connection values.
Basic specification, Position 1, 2 (Approval) = BA, BB, IA, IB
Ex ia
• Power supply and signal circuit with protection type: intrinsic safety Ex ia IIC or IIB.
Service interface (CDI)
Taking the following values into consideration, the device can be connected to the certified Endress+Hauser
FXA291 service tool or a similar interface:
Basic specification, Position 3 (Power Supply; Output) = A TRC [21]
Terminal 1 (+), 2 (–)
Power supply:
U
i
= 30 V
I
i
= 300 mA
P
i
= 1 W
effective inner inductance L
i
= 0 μH
effective inner capacitance C
i
= 12 nF
Basic specification, Position 3 (Power Supply; Output) = B TRC [02]
Terminal 1 (+), 2 (–) Terminal 3 (+), 4 (–)
Power supply: Switch output (PFS):
U
i
= 30 V
I
i
= 300 mA
P
i
= 1 W
U
i
= 30 V
I
i
= 300 mA
P
i
= 1 W
effective inner inductance L
i
= 0 μH
effective inner capacitance C
i
= 5 nF
effective inner inductance L
i
= 0 μH
effective inner capacitance C
i
= 6 nF
Basic specification, Position 3 (Power Supply; Output) = C TRC [04]
Terminal 1 (+), 2 (–) Terminal 3 (+), 4 (–)
Power supply: Output 4...20 mA:
U
i
= 30 V
I
i
= 300 mA
P
i
= 1 W
U
i
= 30 V
I
i
= 300 mA
P
i
= 1 W
effective inner inductance L
i
= 0 μH
effective inner capacitance C
i
= 30 nF
effective inner inductance L
i
= 0 μH
effective inner capacitance C
i
= 30 nF
Service interface
U
i
= 7.3 V
effective inner inductance L
i
= negligible
effective inner capacitance C
i
= negligible
U
o
= 7.3 V
I
o
= 100 mA
P
o
= 160 mW
L
o
(mH) =
C
o
(μF) =
5.00
0.73
2.00
1.20
1.00
1.60
0.50
2.00
0.20
2.60
0.10
3.20
0.05
4.00
0.02
5.50
0.01
7.30
0.005
10.00
0.002
12.70
0.001
12.70