XA00677F-A Micropilot FMR50/51/52/53/54/56/57
34 Endress+Hauser
Position 3 (Hilfsenergie; Ausgang / Power Supply; Output / Alimentation; Sortie) = C (TRC [04])
Position 5 (Gehäuse / Housing / Boîtier) = A
Position 3
(Hilfsenergie; Ausgang /
Power Supply; Output /
Alimentation; Sortie) =
Temperaturklasse /
Temperature class /
Classe de température
P 1 P 2 P 3 P 4 P 5 P 6
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
1 Kanal benutzt /
1 canal used /
1 voie utilisée
T6 ( 85 °C) –40 °C 60 °C 60 °C 60 °C 80 °C 55 °C 80 °C –40 °C –40 °C –40 °C - -
2 Kanäle benutzt /
2 canal used /
2 voies utilisées
T6 ( 85 °C) –40 °C 53 °C 53 °C 53 °C 80 °C 48 °C 80 °C –40 °C –40 °C –40 °C - -
Position 5 (Gehäuse / Housing / Boîtier) = C
Position 3
(Hilfsenergie; Ausgang /
Power Supply; Output /
Alimentation; Sortie) =
Temperaturklasse /
Temperature class /
Classe de température
P 1 P 2 P 3 P 4 P 5 P 6
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
1 Kanal benutzt /
1 canal used /
1 voie utilisée
T6 ( 85 °C) –40 °C 60 °C 60 °C 60 °C 80 °C 58 °C 80 °C –40 °C –40 °C –40 °C - -
2 Kanäle benutzt /
2 canal used /
2 voies utilisées
T6 ( 85 °C) –40 °C 54 °C 54 °C 54 °C 80 °C 52 °C 80 °C –40 °C –40 °C –40 °C - -