Micropilot FMR67B PROFINET with Ethernet-APL
42 Endress+Hauser
Drip-off antenna, PTFE, 50 mm (2 in), with UNI flange and purge air connection
1
R
L2
L1
!42.5 (1.67)
47 (1.85)
A0046488
47 Dimensions of Drip-off antenna, PTFE, 50 mm (2 in), with UNI flange and purge air connection. Unit of
measurement mm (in)
1 Bottom edge of housing
R Reference point of measurement
L1 175 mm (6.89 in); version with Ex d or XP approval +5 mm (+0.20 in)
L2 108 mm (4.25 in); version with Ex d or XP approval +5 mm (+0.20 in)
UNI flange 3"/DN80/80A for Drip-off antenna, PTFE, 50 mm (2 in) with purge air connection
45°
150 (5.9)
160 (6.3)
200 (7.9)
19 (0.75)
M80×1.5
20 (0.79)
A0048875
48 UNI flange 3"/DN80/80A for Drip-off antenna, PTFE, 50 mm (2 in) with purge air connection. Unit of
measurement mm (in)
UNI flange 3"/DN80/80A
Suitable for flange: 3" 150lbs; DN80 PN16; 10K 80A
• Material:
PP
• Process pressure:
≤1 bar (14.5 psi)
• Weight:
0.5 kg (1.10 lb)