Technical data Micropilot FMR67B PROFINET with Ethernet-APL
86 Endress+Hauser
Process connection UNI flange, aluminum adjustable
Seal T
p
Process pressure range
A0048812
Graphite –40 to +280 °C (–40 to +536 °F) –1 to 1 bar (–14.5 to 14.5 psi)
The pressure range may be further restricted in the event of a CRN approval.
Antenna flush mount, PTFE, 80 mm (3 in)
Process connection standard flange with purge air connection
Seal T
p
Process pressure range
A0047828
FKM Viton GLT –40 to +150 °C (–40 to +302 °F) –1 to 16 bar (–14.5 to 232 psi)
FKM Viton GLT –40 to +200 °C (–40 to +392 °F) –1 to 16 bar (–14.5 to 232 psi)
Process connection UNI flange 316L with purge air connection
Seal T
p
Process pressure range
A0047829
FKM Viton GLT –40 to +150 °C (–40 to +302 °F) p
gauge
= –1 to 1 bar (–14.5 to 14.5 psi)
FKM Viton GLT –40 to +200 °C (–40 to +392 °F) p
gauge
= –1 to 1 bar (–14.5 to 14.5 psi)
Process connection UNI flange, aluminum adjustable with purge air connection
Seal T
p
Process pressure range
A0047830
FKM Viton GLT –40 to +150 °C (–40 to +302 °F) –1 to 1 bar (–14.5 to 14.5 psi)
FKM Viton GLT –40 to +200 °C (–40 to +392 °F) –1 to 1 bar (–14.5 to 14.5 psi)
The pressure range may be further restricted in the event of a CRN approval.
Dielectric constant For bulk solids
ε
r
≥ 1.6
Contact Endress+Hauser for applications with lower dielectric constants than indicated.