Operating menu Tankside Monitor NRF81
158 Endress+Hauser
User entry 0 to 255
Factory setting 1
Additional information
Read access Operator
Write access Maintenance
Level mapping
Navigation  Setup → Advanced setup → Communication → V1 X1-4 → Configuration → Level
mapping
Prerequisite Communication interface protocol (→  153) = V1
Description Determines the transmittable range of levels.
Selection +ve
+ve & -ve
Factory setting +ve
Additional information
Read access Operator
Write access Maintenance
In V1, the level is always represented by a number in the range from 0 to 999 999. This
number corresponds to a level as follows:
"Level mapping" = "+ve"
Number Corresponding level
0 0.0 mm
999 999 99 999.9 mm
"Level mapping" = "+ve & -ve"
Number Corresponding level
0 0.0 mm
500 000 50 000.0 mm
500 001 –0.1 mm
999 999 –49 999.9 mm