Operating menu Tankside Monitor NRF81
244 Endress+Hauser
15.4.3 "Simulation" submenu
Read access Maintenance
Navigation Diagnostics → Simulation
Device alarm simulation
Navigation  Diagnostics → Simulation → Device alarm simulation
Description Switch the device alarm on and off.
Selection Off
On
Factory setting Off
Additional information
Read access Operator
Write access Maintenance
Diagnostic event simulation
Navigation  Diagnostics → Simulation → Diagnostic event simulation
Description Select a diagnostic event to simulate this event.
Selection The diagnostic events of the device
Factory setting Off
Additional information
Read access Operator
Write access Maintenance
To terminate the simulation, select Off.
Current output simulation
Navigation  Diagnostics → Simulation → Current output 1 simulation
 Diagnostics → Simulation → Current output 2 simulation
Prerequisite The device has an Anlog I/O module.
Operating mode (→  138) = 4..20mA output or HART slave +4..20mA output